Anecdotal and documented reports from both patients and doctors have described unanticipated breaks in connections between implanted catheters and drug pumps. In extreme cases, such disconnections in patient-required therapies could result in either withdrawal symptoms or possible deaths. Patients typically attribute such device failures to falls or impacts associated with vigorous physical activity; subsequent failure analyses most often have indicated pump connector uncouplings. We fabricated a facsimile of the II pump that included both an accelerometer and a force sensor. The force sensor measured forces imparted on the pump connector via the attached catheter and surrounding tissues. The test pump was implanted in the lower left abdominal areas of porcine cadavers in various orientations. Wire-reinforced catheters were tunneled for under the abdominal epidermis, anteriorly toward the head, and the non-connector pump ends were secured by sutures. Following each simulated implant, the cadaver specimens were loaded into a harness and hoisted to a height where either their buttocks or backs were above the floor, simulating a worst-case scenario in which a patient might have fallen down a flight of stairs or off a step stool. The cadavers were then quick released from the hoist attachment, while forces (, , and ) and accelerations (, , and ) versus time were simultaneously recorded. Six porcine cadaver specimens were utilized for a total of 72 trials. Subsequent Monte Carlo analyses allowed us to model the variation in stress imparted onto the pump connectors and the estimated variation of the pump connector strength, as a means of predicting required connector retention impact specification for future designs. The recorded forces applied onto the connectors, including data from all three connector axes (, , and ), were typically within the range of . However, in several trials, applied forces ranged as high as . Monte Carlo modeling provided a maximum resultant load specification of for a duration. Based on this value, due to predicted impact events, subsequent failures of future designs would be estimated at . Based on our data, a new design requirement has been generated to ensure that implantable drug pump connector assemblies will, in high probability, perform their intended functions.
Skip Nav Destination
e-mail: thomas.c.bischoff@medtronic.com
e-mail: marty.d.martens@medtronic.com
e-mail: matthew.adams@medtronic.com
e-mail: galla009@umn.edu
e-mail: iaizz001@umn.edu
Article navigation
June 2007
Research Papers
Measurement of Impact Loads Applied to an Implanted Drug Pump Connector in a Porcine Cadaver Specimen
Thomas C. Bischoff,
e-mail: thomas.c.bischoff@medtronic.com
Thomas C. Bischoff
Medtronic Neurological
, 800 53rd Avenue NE, MS N363, Columbia Heights, MN 55421
Search for other works by this author on:
Marty D. Martens,
e-mail: marty.d.martens@medtronic.com
Marty D. Martens
Medtronic Neurological
, 800 53rd Avenue NE, MS N363, Columbia Heights, MN 55421
Search for other works by this author on:
Matthew H. Adams,
e-mail: matthew.adams@medtronic.com
Matthew H. Adams
Medtronic Neurological
, 800 53rd Avenue NE, MS N363, Columbia Heights, MN 55421
Search for other works by this author on:
William J. Gallagher,
William J. Gallagher
Department of Surgery,
e-mail: galla009@umn.edu
University of Minnesota
, B172 Mayo, MMC 107, 420 Delaware Street SE, Minneapolis, MN 55455
Search for other works by this author on:
Paul A. Iaizzo
Paul A. Iaizzo
Departments of Surgery and Physiology,
e-mail: iaizz001@umn.edu
University of Minnesota
, B172 Mayo, MMC 107, 420 Delaware Street SE, Minneapolis, MN 55455
Search for other works by this author on:
Thomas C. Bischoff
Medtronic Neurological
, 800 53rd Avenue NE, MS N363, Columbia Heights, MN 55421e-mail: thomas.c.bischoff@medtronic.com
Marty D. Martens
Medtronic Neurological
, 800 53rd Avenue NE, MS N363, Columbia Heights, MN 55421e-mail: marty.d.martens@medtronic.com
Matthew H. Adams
Medtronic Neurological
, 800 53rd Avenue NE, MS N363, Columbia Heights, MN 55421e-mail: matthew.adams@medtronic.com
William J. Gallagher
Department of Surgery,
University of Minnesota
, B172 Mayo, MMC 107, 420 Delaware Street SE, Minneapolis, MN 55455e-mail: galla009@umn.edu
Paul A. Iaizzo
Departments of Surgery and Physiology,
University of Minnesota
, B172 Mayo, MMC 107, 420 Delaware Street SE, Minneapolis, MN 55455e-mail: iaizz001@umn.edu
J. Med. Devices. Jun 2007, 1(2): 119-125 (7 pages)
Published Online: March 20, 2007
Article history
Received:
April 1, 2006
Revised:
March 20, 2007
Citation
Bischoff, T. C., Martens, M. D., Adams, M. H., Gallagher, W. J., and Iaizzo, P. A. (March 20, 2007). "Measurement of Impact Loads Applied to an Implanted Drug Pump Connector in a Porcine Cadaver Specimen." ASME. J. Med. Devices. June 2007; 1(2): 119–125. https://doi.org/10.1115/1.2736398
Download citation file:
31
Views
Get Email Alerts
Cited By
Related Articles
A Novel Combination Therapy for Post-Operative Arrhythmias
J. Med. Devices (June,2009)
CT Visualization of Cryoablation in Pulmonary Veins
J. Med. Devices (June,2009)
Design of an Autoclavable Active Cannula Deployment Device
J. Med. Devices (June,2011)
Related Proceedings Papers
Related Chapters
Process Components
Bioprocessing Piping and Equipment Design: A Companion Guide for the ASME BPE Standard
mDFA Human Empirical Results
Modified Detrended Fluctuation Analysis (mDFA)
Experimental Studies
Nanoparticles and Brain Tumor Treatment