Background: While it is established that mechanical heart valves (MHVs) damage blood elements during leakage and forward flow, the role in thrombus formation of platelet activation by high shear flow geometries remains unclear. In this study, continuously recalcified blood was used to measure the effects of blood flow through orifices, which model MHVs, on the generation of procoagulant thrombin and the resulting formation of thrombus. The contribution of platelets to this process was also assessed. Method of Approach: 200, 400, 800, and orifices simulated the hinge region of bileaflet MHVs, and 200, 400, and wide slits modeled the centerline where the two leaflets meet when the MHV is closed. To assess activation of coagulation during blood recirculation, samples were withdrawn over and the plasmas assayed for thrombin-antithrombin-III (TAT) levels. Model geometries were also inspected visually. Results: The 200 and round orifices induced significant TAT generation and thrombosis over the study interval. In contrast, thrombin generation by the slit orifices, and by the 800 and round orifices, was negligible. In additional experiments with nonrecalcified or platelet-depleted blood, TAT levels were markedly reduced versus the studies with fully anticoagulated whole blood . Conclusions: Using the present method, a significant increase in TAT concentration was found for 200 and orifices, but not 800 and orifices, indicating that these flow geometries exhibit a critical threshold for activation of coagulation and resulting formation of thrombus. Markedly lower TAT levels were produced in studies with platelet-depleted blood, documenting a key role for platelets in the thrombotic process.
Skip Nav Destination
e-mail: ajit.yoganathan@bme.gatech.edu
Article navigation
February 2006
Technical Papers
Flow and Thrombosis at Orifices Simulating Mechanical Heart Valve Leakage Regions
Anna M. Fallon,
Anna M. Fallon
School of Chemical Engineering,
Georgia Institute of Technology
, Atlanta, GA 30332
Search for other works by this author on:
Nisha Shah,
Nisha Shah
School of Chemical Engineering,
Georgia Institute of Technology
, Atlanta, GA 30332
Search for other works by this author on:
Ulla M. Marzec,
Ulla M. Marzec
Department of Biomedical Engineering,
Oregon Health and Science University
, Portland, OR 97006
Search for other works by this author on:
James N. Warnock,
James N. Warnock
Woodruff School of Mechanical Engineering,
Georgia Institute of Technology
, Atlanta, GA 30332
Search for other works by this author on:
Ajit P. Yoganathan,
Ajit P. Yoganathan
Wallace H. Coulter School of Biomedical Engineering,
e-mail: ajit.yoganathan@bme.gatech.edu
Georgia Institute of Technology and Emory University
, Atlanta, GA 30332
Search for other works by this author on:
Stephen R. Hanson
Stephen R. Hanson
Department of Biomedical Engineering,
Oregon Health and Science University
, Portland, OR 97006
Search for other works by this author on:
Anna M. Fallon
School of Chemical Engineering,
Georgia Institute of Technology
, Atlanta, GA 30332
Nisha Shah
School of Chemical Engineering,
Georgia Institute of Technology
, Atlanta, GA 30332
Ulla M. Marzec
Department of Biomedical Engineering,
Oregon Health and Science University
, Portland, OR 97006
James N. Warnock
Woodruff School of Mechanical Engineering,
Georgia Institute of Technology
, Atlanta, GA 30332
Ajit P. Yoganathan
Wallace H. Coulter School of Biomedical Engineering,
Georgia Institute of Technology and Emory University
, Atlanta, GA 30332e-mail: ajit.yoganathan@bme.gatech.edu
Stephen R. Hanson
Department of Biomedical Engineering,
Oregon Health and Science University
, Portland, OR 97006J Biomech Eng. Feb 2006, 128(1): 30-39 (10 pages)
Published Online: August 22, 2005
Article history
Received:
June 28, 2004
Revised:
August 22, 2005
Citation
Fallon, A. M., Shah, N., Marzec, U. M., Warnock, J. N., Yoganathan, A. P., and Hanson, S. R. (August 22, 2005). "Flow and Thrombosis at Orifices Simulating Mechanical Heart Valve Leakage Regions." ASME. J Biomech Eng. February 2006; 128(1): 30–39. https://doi.org/10.1115/1.2133768
Download citation file:
Get Email Alerts
Analysis of Transient Cutting Forces in Cortical Bone During Ultrasonically Assisted Cutting
J Biomech Eng (June 2025)
Related Articles
The Osmotic Swelling Characteristics of Cardiac Valve Prostheses
J Biomech Eng (August,2000)
Numerical Study of Shear-Induced Thrombus Formation Over Aterial Stent Struts
J. Med. Devices (June,2009)
Two-Dimensional Simulation of Flow and Platelet Dynamics in the Hinge Region of a Mechanical Heart Valve
J Biomech Eng (March,2009)
Vortex Shedding as a Mechanism for Free Emboli Formation in Mechanical Heart Valves
J Biomech Eng (April,2000)
Related Proceedings Papers
Related Chapters
Introduction
Mechanical Blood Trauma in Circulatory-Assist Devices
Concluding remarks
Mechanical Blood Trauma in Circulatory-Assist Devices
Introduction
Modified Detrended Fluctuation Analysis (mDFA)